粉体行业在线展览
面议
1940
--
The gas analyser MCA 04 can be used in
emission measuring facilities as well as in systems of process and
analysis measuring technology. The MCA 04 is approved for measurements
according to 13th, 17th and 27th BlmSchV and TI Air.
partial view - pivoting frame with analyser and FID
By means of the gas analyser MCA 04 up to 7 gas components can be
measured at the same time - as e.g. CO (carbon monoxide), NO (nitrogen
oxide), NO2 (nitrogen dioxide), SO2 (sulphur dioxide), HCl (hydrogen
chloride), NH3 (ammonium hydroxide), N2O (nitrous oxide, laughing gas) ,
H2O (water, humidity), CO2 (carbon dioxide) and O2 (oxygen).
Technical data
Case: 19""-rack, 4 HU, dimensions 665 x 440 x 360 mm (W x H x D), ca. 40 kg, IP 52
Measuring principle: CO, NO, NO2, SO2, HCl, NH3, H2O, CO2, N2O: infrared
absorption (single-beam dual-wavelength method respectively gas filter
correlation)
O2: ZrO2-cell
Smallest measuring ranges: CO: 0 ... 75 mg/m3
NO: 0 ... 200 mg/m3
NO2: 0 ... 50 mg/m3
SO2: 0 ... 75 mg/m3
HCl: 0 ... 15 mg/m3
NH3: 0 ... 30 mg/m3
H2O: 0 ... 40 Vol%
CO2: 0 ... 20 Vol%
N2O: 0 ... 50 mg/m³
O2: 0 ... 25 Vol%
Calibration: zero point: automatically with ambient air; sensitivity: with test gas
Display: 7.4"" black/white LC-Display (640*480 Pixel)
Media temperature: max. 200 °C
Ambient temperature: +5 ... +35 °C
Interfaces: RS 232, optionally MOD-Bus
Analogue signals: max. 8 x 4 ... 20 mA
Digital signals: D/O: for failure, maintenance, maintenance request and measuring range signalisation available (230 V, 1 A)
Digital inputs: optionally for analogue and digital signals
Power supply: 110 VAC, 230 VAC
TÜV-approval: TI air, 13th, 17th and 30th BImSchV
LHTG/LHTM/LHTW
Empyrean
V-Sorb4800-金埃谱
EMIA-820V
Hydrolink
Autoflex R837
3H-2000A
SQL810C/1010C
UNI800B
略
BI-ZTU